Replacing Fuses/Changing Line Voltage DMM6500 DAQ6510

Question :

Replacing Fuses/Changing Line Voltage DMM6500 DAQ6510

Answer :

  • Line Voltage
    • Remove the fuse housing above the line power plug.
    • Remove the fuse from the housing and replace with the appropriate, functioning fuse.
    • Change the power line voltage¬†setting by removing the inner housing for the fuse and rotating the block to display the appropriate setting through the window of the main fuse housing front panel.
    • Insert back into the housing and insert the replacement fuse into the complete housing.
    • Insert the housing back into the slot on the instrument until an audible click and the housing is secured.
  • High Power Fuse
    • Remove the Communication Accessory Card or slot cover.
    • Pull up on the tab of the fuse cover to access the high power (10A and 1000V) fuse for replacement.
    • Return the fuse cover and accessory card/slot cover.
  • Current Fuse
    • A current fuse can be found in both the front and rear panels of the instrument.
    • Press on the white input jack and rotate the housing counter-clockwise 90 degrees.
    • Release and remove housing from the instrument.
    • Replace the fuse and return the housing to the instrument, rotating 90 degrees clockwise.

FAQ ID 714431