My trigger is set to "trigger immidately" or "If anything then trigger". Why is my memory only half full?

Question :

My trigger is set to "trigger immidately" or "If anything then trigger". Why is my memory only half full?

Answer :

This is because the default trigger position is set to 50% memory before the trigger and 50% memory after the trigger. Since the Logic Analyzer is triggering off the first thing it sees, it does not get a chance to fill up the first 50%. To fix this, go to the trigger setup menu and change the trigger position to 0% or check the "force main prefil" box or if this is unavailable (some models) then try sliding the "trigger position" slider as far as you can to the left.

This FAQ Applies to:

Product Series: TLA7000

Product: